A method for free thyroxine measurement in human serum using equilibrium dialysis followed by liquid chromatography tandem mass spectrometry (LC-MS/MS) is described. Free thyroxine in serum is first separated from protein-bound thyroxine by equilibrium dialysis and then measured by LC-MS/MS.
Keywords: Equilibrium dialysis; Free thyroxine; LC-MS/MS; T4.
© 2022. The Author(s), under exclusive license to Springer Science+Business Media, LLC, part of Springer Nature.